Ultra-Low Refrigerated-Heating Circulators
For working temperatures from -91 °C to +200 °C
JULABO Ultra-Low Refrigerated Circulators for heating and cooling. The units are suitable for external temperature applications with simultaneous operation in the bath tank. The units provide particularly high heating and cooling capacities for rapid heat-up and cool-down times even for large volume, external systems. ‚FP‘ models offer proportional cooling control to ensure energy-saving performance and to reduce heat generation to the environment. Models with the designation ‚W‘ are water-cooled. With handle or castors to allow easy transport. With drain port for conveniently draining the bath fluid. The units offer an improved insulation to prevent ice-formation and have a filling level display. Typical applications: Temperature applications to jacketed reaction vessels, autoclaves, miniplant installations, kilo labs, freezing point determination, calibration at low temperatures, petroleum testing, etc.
Upper middle class with ext. Pt100 sensor connection
Ultra-low Refrigerated Circulator with 2-stage cascade-cooling
JULABO Ultra-Low Refrigerated Circulators for heating and cooling are suitable for external temperature tasks. Powerful circulating pump systems and high heating and cooling capacities guarantee short heat-up and cool-down times. ACC ‘Active Cooling Control‘ provides active cooling control across the whole temperature range. An energy saving feature with only minor loss of heat is provided by the proportional cooling capacity control in the ‘FP’ refrigeration units. The inevitable dust accumulation that occurs in refrigeration systems is solved by means of an easily removed venting grill, which allows for easy cleaning. The units are equipped with handles or castors for an easy transportation. All models have a drain tap on the front allowing the fluid to be drained easily. Additionally the instruments are equipped with an improved insulation to help avoid ice-formation, and have a visual liquid level display.
Your advantages
- VFD COMFORT DISPLAY
- Keypad for setpoints, warning/safety values and menu functions
- PID3 cascade temperature control
- ATC3 3-Point-Calibration
- Pt100 External sensor connection for measurement and control
- SMART PUMP, electronically adjustable pump stages
- Early warning system for low liquid level (DBGM 203 06 059.8)
- Adjustable high temperature cut-out, visible via display
- RS232 interface for online communication
- Integrated programmer for 10 program steps
- Active Cooling Control
Order No. | Model | Working temperature range | Temperature stability | Cooling capacity at 20°C (kW) | Heating capacity (kW) | Pump capacity flow rate | Pump pressure |
(°C) | (°C) | (l/min) | (bar) | ||||
9162670 | F70-ME | -70 … +100 | ±0.02 | 0.34 | 1.3 | 11-16 | 0.23-0.45 |
9162681 | F81-ME | -81 … +100 | ±0.02 | 0.45 | 1.3 | 11-16 | 0.23-0.45 |
9162689 | FP89-ME | -90 … +100 | ±0.02 | 1 | 1.3 | 11-16 | 0.23-0.45 |
Superior models for most demanding applications
JULABO Ultra-Low Refrigerated Circulators for heating and cooling are suitable for external temperature tasks. Powerful circulating pump systems and high heating and cooling capacities guarantee short heat-up and cool-down times. ACC ‘Active Cooling Control‘ provides active cooling control across the whole temperature range. An energy saving feature with only minor loss of heat is provided by the proportional cooling capacity control in the ‘FP’ refrigeration units. The inevitable dust accumulation that occurs in refrigeration systems is solved by means of an easily removed venting grill, which allows for easy cleaning. The units are equipped with handles or castors for an easy transportation. All models have a drain tap on the front allowing the fluid to be drained easily. Additionally the instruments are equipped with an improved insulation to help avoid ice-formation, and have a visual liquid level display.
Your advantages
- VFD COMFORT DISPLAY
- LCD DIALOG DISPLAY backlit for convenient interactive operation
- Keypad for setpoints, warning/safety values and menu functions
- ICC (Intelligent Cascade Control), self-optimizing temperature control
- TCF Temperature Control Features to optimize the control behaviour
- ATC3 3-Point-Calibration
- Pt100 External sensor connection for measurement and control
- SMART PUMP, electronically adjustable pump stages
- Early warning system for low liquid level (DBGM 203 06 059.8)
- Adjustable high temperature cut-out, visible via display
- RS232/RS485 interface for online communication
- Integrated programmer for 6 x 60 program steps
- Connections for solenoid valve and HSP booster pump
- Proportional cooling control
- Active Cooling Control
- Proportional cooling control
Order No. | Model | Working temperature range | Temperature stability | Cooling capacity at 20°C (kW) | Heating capacity (kW) | Pump capacity flow rate | Pump pressure |
(°C) | (°C) | (l/min) | (bar) | ||||
9312681 | F81-HL | -81 … +100 | ±0.02 | 0.45 | 1.3 | 22-26 | 0.4-0.7 |
9352795N | F95-SL | -95 … +0 | ±0.05 | 3 | 22-26 | 0.4-0.7 | |
9352751 | FP51-SL | -51 … +200 | ±0.05 | 2 | 3 | 22-26 | 0.4-0.7 |
9352752 | FP52-SL | -60 … +100 | ±0.05 | 3 | 3 | 22-26 | 0.4-0.7 |
9352752N | FP52-SL | -60 … +100 | ±0.05 | 3 | 3 | 22-26 | 0.4-0.7 |
9352752N150 | FP52-SL-150C | -60 … +150 | ±0.05 | 3 | 3 | 22-26 | 0.4-0.7 |
9352755 | FP55-SL | -60 … +100 | ±0.05 | 5.2 | 3 | 22-26 | 0.4-0.7 |
9352755N | FP55-SL | -60 … +100 | ±0.05 | 5.2 | 3 | 22-26 | 0.4-0.7 |
9352755N150 | FP55-SL-150C | -60 … +150 | ±0.05 | 5.2 | 3 | 22-26 | 0.4-0.7 |
9312689 | FP89-HL | -90 … +100 | ±0.02 | 1 | 1.3 | 22-26 | 0.4-0.7 |
9352790 | FP90-SL | -90 … +100 | ±0.05 | 1.8 | 3 | 22-26 | 0.4-0.7 |
9352790N | FP90-SL | -90 … +100 | ±0.05 | 1.8 | 3 | 22-26 | 0.4-0.7 |
9352790N150 | FP90-SL-150C | -90 … +150 | ±0.05 | 1.8 | 3 | 22-26 | 0.4-0.7 |
9352753 | FPW52-SL | -60 … +100 | ±0.05 | 3 | 3 | 22-26 | 0.4-0.7 |
9352753N | FPW52-SL | -60 … +100 | ±0.05 | 3 | 3 | 22-26 | 0.4-0.7 |
9352753N150 | FPW52-SL-150C | -60 … +150 | ±0.05 | 3 | 3 | 22-26 | 0.4-0.7 |
9352756 | FPW55-SL | -60 … +100 | ±0.05 | 5.5 | 3 | 22-26 | 0.4-0.7 |
9352756N | FPW55-SL | -60 … +100 | ±0.05 | 5.5 | 3 | 22-26 | 0.4-0.7 |
9352756N150 | FPW55-SL-150C | -60 … +150 | ±0.05 | 5.5 | 3 | 22-26 | 0.4-0.7 |
9352791 | FPW90-SL | -90 … +100 | ±0.05 | 1.8 | 3 | 22-26 | 0.4-0.7 |
9352791N | FPW90-SL | -90 … +100 | ±0.05 | 1.8 | 3 | 22-26 | 0.4-0.7 |
9352791N150 | FPW90-SL-150C | -90 … +150 | ±0.05 | 1.8 | 3 | 22-26 | 0.4-0.7 |
9352793 | FPW91-SL | -91 … +100 | ±0.2 | 4.5 | 3 | 22-26 | 0.4-0.7 |
9352793N | FPW91-SL | -91 … +100 | ±0.2 | 4.5 | 3 | 22-26 | 0.4-0.7 |
9352796N | FW95-SL | -95 … +0 | ±0.05 | 3 | 22-26 | 0.4-0.7 |